DAVIS 종합 계측기 - 한국대리점 코린스계기(주)
|
UM68900 Multi-Calibrator for process signals |
|
|
|
 | |
Features |
- Sources and reads millivolts, volts and milliamps
- Powers, reads and simulates 2-wire transmitters
- Simultaneous input and output readings
- Fixed calibration steps
- Automatic signal ramping
|
| |
|
Calibrator includes |
- Carrying case
- Batteries
- Instruction manual
- NIST Certification available
- Spare fuse
- Set of test leads
|
| | SPECIFICATIONS
Function |
Range |
Resolution |
Accuracy % of reading |
Remarks |
Measure mVolts |
0-120mV |
0.01 mV |
±0.017% ±1 digit |
R-input ≥20 MÆ |
Output mVolts |
0-120mV |
0.01mV |
±0.017% ±1 digit |
R-output 0,2Æ |
Measure Volts |
0-120V |
0.01V |
±0.04% ±1 digit |
R-input 1 MÆ |
Output Volts |
0-12V |
0.001V |
±0.017% ±1 digit |
R-output 0,2 Æ |
Measure mA |
0-52mA |
0.01mA |
±0.04% ±1 digit |
R-input 3,5 Æ fused |
Output mA |
0-24mA |
0.01mA |
±0.04% ±1 digit |
R-max. 900 Æ |
Transmitter Sim. |
0-24mA |
0.01mA |
±0.04% ±1 digit |
V-max ext. 56 Volts |
Special Functions |
1. Fixed steps (programmable, 10% or 20% divisions) |
2. Signal ramping (up/dwell/down)Reference 22°C ±1 |
Calibration |
Traceable to National Standards with correlation to NIST |
Long Term Stability |
0.02%/year of span |
Indicated Accuracies |
Specified for 15°C to 35°C. Outside these limits; ± 1/sd max. on zero and ±0.001%/°C of span |
Operating Temperature |
-10°C to +50°C (14°F to 122°F) |
Storage Temperature |
-20°C to +70°C (-4°F to 158°F) |
Warm up Time |
2 minutes to rated accuracy |
Relative Humidity |
0 to 90% non-condensing |
Input/output Protection |
70 volts on millivolt input, 30Volts on Volts input |
Input/output Isolation |
500VDC continuous |
Output Voltage |
Protected by load monitor, displays "check loop" |
Output Current |
Protected by load monitor, display "check loop" |
Transmitter Power Supply |
Available for user; 24VDC current limited at 34mA |
Read-out |
Simultaneous read-out of input and output values. Dialogue in english |
Batteries |
4 x 1.5Volt, type R14 (Size C alkaline) or rechargeable NiCad |
Battery Life |
22 hours with alkaline batteries at 20°C (68°F), 7 hours with 20mA load |
Low-battery Indication |
Pre-warning alternately flashes "bat" |
External Power Supply |
By optional charge/line adapter p/n 13603 230/115V—50/60 Hz |
Connections |
Suitable for 4mm test plugs |
Protection |
IP 53 |
Housing |
Textured high-impact ABS plastic |
Size |
200 x 117 x 32mm (8.0" x 4.7" x 1.3") without carrying case |
Weight |
0.9 kg ( 2 pounds) including batteries, carrying case and test leads |
CAT NO. |
 |
MODEL |
 |
DEscRIPTION |
 |
AVAIL |
 |
PRICE |
UM68900 |
|
MVX |
|
MULTI-CALIBRATOR FOR PROCESS SIGNALS (mV,V,mA) |
|
Yes |
|
2,383,128 |
UM13603 |
|
191-A005 |
|
OPTIONAL Power Supply, 110V FOR Process Signal Multi-Calibrator |
|
Yes |
|
201,960
|
UM063 |
|
191-A010 |
|
OPTIONAL SET OF TEST LEADS FOR MVX FOR Process Signal Multi-Calibrator |
|
Yes |
|
73,440
| | |
|
|
 | |
 |
|
 |
|
이 상품에 대한 사용후기가 아직 없습니다.
사용후기를 작성해 주시면 다른 분들께 많은 도움이 됩니다. |
|
 |
|
 |
|
|
 | |
 |
|
 |
|
이 상품에 대한 질문이 아직 없습니다.
궁금하신 사항은 이곳에 질문하여 주십시오. |
|
 |
|
 |
|
|
|
DAVIS 종합 계측기 - 한국대리점 코린스계기(주)
|
|
|
회사소개 | 서비스이용약관 | 개인정보취급방침 |
주소 : 경기도 성남시 중원구 둔촌대로 474, 708호 (상대원동 선텍시티1)
Tel : 031-777-1588 / Fax : 031-777-1587 / 개인정보관리책임자 : 관리자
사업자 등록번호 : 116-81-31753 / 통신판매업신고 : 2005-경기성남-0886호
대표자 : 홍영화 / 운영자 : 윤상미
상호 : 주식회사 코린스 / Copyright ⓒ 2002. All Rights Reserved.
|
|
|
|